20000004342IKA® RC 2 basicRC 2_112014Reg.-No. 4343-01Operating instructions EN
10Before commissioning, make sure that the device has not been moved for one hour.Check whether the voltage specied on the type plate matc
11Menu settings• Maximum temperature (HIT):The maximum adjustable value: 40 °C.• Minimum temperature (LO T):The minimum adjustable value: -20 °C.• Max
12• Temperature control type (AUTO):AUTO 1: AUTO 1 is the default setting. The default PID settings are used automatically.Cooling down curve in AUTO
13Set the rst point temperature (step 1, e.g. 10 °C) with “Pump (+)” button (F) or “Pump (-)” button (G). Con-rm the setting by pressing “OK/Pump“ b
14Command syntax and format:The following applies to the command set:- Commands are generally sent from the computer (Master) to the device (Slave).-
15USB cable A - Micro B 2.0:This cable is required to connect USB port (11) to a PC.USB micro BUSB AAFig. 15Maintenance and cleaningTo avoid contamina
16Error codesAccessories• Water bath protective media:AQ.1 Aquacide 1• Tubing, hoses, adapters and couplings:LT 5.20 Metal hose (isolated M16 x 1)
17Technical dataNominal voltage VAC 230 ± 10% / 115 ± 10% Frequency Hz 50 / 60 Max. input power W 300Working temperature range °C -20 ... RTOperating
18WarrantyIn accordance with IKA® warranty conditions, the warranty period is 24 months. For claims under the warranty please contact your local deale
IKA® - Werke GmbH & Co.KGJanke & Kunkel-Str. 10D-79219 StaufenTel. +49 7633 831-0Fax +49 7633 [email protected]
2Fig. 11451212Fig. 2237681011121249Item Designation 1 Operator panel and display2 Mains switch3 Power socket 4 Ventin
3ContentsEN PageEC Declaration of Conformity ...
4Explication of warning symbolsSafety instructionsGeneral information:• Read the operating instructions completely before starting up and follow the s
5• A laboratory circulator refrigerates and circulates uid according to specied parameters. This involves hazards due to low temperatures and ge
6Unpacking• Unpacking:- Please unpack the device carefully- In the case of any damage a detailed report must be sent immediately (post, rail or forwar
7Fig. 6• Filling and draining:- Before lling the uid into the bath, open the venting grid as indicated in following gure.- Check and make sure th
8• Fluids (Standard information for IKA® uid):No.IKA® Designation Operating temperature range for open bath application(°C)Operating temperature rang
9max. 10 ˚• Moving the device:Empty all fluid in the bath before moving device from one place to other place.The device can be lifted up and moved by