8
• Fluids (Standard information for IKA
®
uid):
No.
IKA
®
Designation
Operating temperature
range for open bath
application
(°C)
Operating temperature
range for closed bath
applications
(°C)
Safety temperature
(°C)
Flash point
(°C)
0
CF.EG28.N10.80.8
-10 ... 80 -10 … 80 80 115
1
CF.EG39.N20.80.16
-20 … 80 -20 … 80 80 115
2
CF.EG44.N25.80.19
-25 … 80 -25 … 80 80 115
3
CF.EG48.N30.80.22
-30 … 80 -30 … 80 80 115
4
UF.Si.N30.150.10LV
-30 … 130 -30 … 150
145
>170
5
HF.Si.20.200.50
20 … 200 20 … 200 255 >280
6
HF.Si.20.250.50A
20 … 200 20 … 250 255 >280
7
Water
5 … 95 5 … 95 - -
8
Customized
The listed fluid temperature range may deviate from the device operating temperature ranges and settings.
Note: for open bath application!
Note: Tap water may be unsuitable for operation because the calcium carbonate content may cause calcification.
High purity water (from icon exchangers) and distilled or bi-distilled water are unsuitable for operation due to corrosive
properties of these media. High purity water and distillates are suitable as a medium after adding 0.1 g soda (Na
2
CO
3
,
sodium carbonate) per liter of water.
Nomenclaturefor IKA
®
fluids:
CF.EG28.N10.80.8 --
(1) Classication:
HF: Heating Fluid
CF: Cooling Fluid
UF: Universal Fluid
(2) Chemical composition:
Si: Silicone oil
EG: Ethylene Glycol
(3) Temperature range: (
Minimum temperature. Maximum temperature)
N: Negative Temperature
(4) Viscosiy:
Viscosiy at 25 °C for Heating Fluid (HF)
Viscosity at -20 °C for Cooling Fluid (CF)
Viscosity at 25 °C for Universal Fluid (UF)
Dynamic viscosity [mPa.s] is a product of kinematic viscosity [mm²/s] and density [kg/m³] of the uid divided by 1000.
(5) Additional information:
A: Oil Additives
LV: Low Viscosity
(5) Addition information
(4) Viscosity (8
mm²/s
)
(3) Temperature range (-10 – 80 °C)
(2) Chemical composition (Ethylene Glycol 28%)
(1) Classication (Cooling fluid)