Command syntax and format:
The following applies to the command set:
- Commands are generally sent from the computer (Master)
to the device (Slave).
- The device sends only at the computer’s request. Even
fault indications cannot be sent spontaneously from the
device to the computer (automation system).
- Commands are transmitted in capital letters.
- Commands and parameters including successive parame-
ters are separated by at least one space (Code: hex 0x20).
- Each individual command (incl. parameters and data)
and each response are terminated with Blank CR LF
(Code: hex 0x20 hex 0x0d hex 0x20 hex 0x0A) and have
a maximum length of 80 characters.
- The decimal separator in a number is a dot (Code: hex
0x2E).
The above details correspond as far as possible to the rec-
ommendations of the NAMUR working party (NAMUR rec-
ommendations for the design of electrical plug connections
for analogue and digital signal transmission on individual
items of laboratory control equipment, rev. 1.1).
The NAMUR commands and the additional specic IKA
®
co mmands serve only as low level commands for commu-
nication between the device and the PC. With a suitable
terminal or communications programme these commands
can be transmitted directly to the circulator equipment. The
IKA
®
software package, Labworldsoft
®
, provides a conve-
nient tool for controlling circulating equipment and collect-
ing data under MS Windows, and includes graphical entry
features, for pump motor speed ramps for example.
PC 1.1 Cable:
This cable is required to connect RS 232 port (10) to a PC.
1
2 RxD
3 TxD
4
5 GND
6
7 RTS
8 CTS
9
1
RxD 2
TxD 3
4
GND 5
6
RTS 7
CTS 8
9
PC
1
2
3
4
5
6
7
8
9
9
8
7
6
5
4
3
2
1
Fig. 14
Commands:
NAMUR
Commands
Function
IN_PV_2 Read the internal actual temperature
IN_PV_4 Read the pump actual speed
IN_SP_1 Read the internal setting temperature
IN_SP_4 Read the pump setting speed
IN_TMODE Read temperature control
0: internal regulation
OUT_SP_1 xxx Set the internal setting temperature XXX
OUT_SP_12@n Set the WD safety temperature with echo of the set (dened) value.
OUT_SP_4 xxx Set the pump speed XXX
OUT_SP_42@n Set the WD-safety speed with echo of the set (dened) value.
OUT_WD1@n Start the watchdog mode 1 and set the time for the watchdog to n (20...1500) seconds.
Echo of the Watchdog time.
During a WD1-event, the heating and pump functions are switched off and message PC 1 is displayed
This command needs to be send within the watchdog time.
OUT_WD2@n Start the watchdog mode 2 and set the watchdog time to n (20...1500) seconds.
Echo of the watchdog time.
During a WD2-event, the set temperature is changed to the WD safety temperature and the pump
set speed is set to the WD safety speed. The PC 2 warning is displayed.
This command needs to be send within the watchdog time.
RESET Reset the PC control and stop the device functions.
START_1 Start the tempering function
START_4 Start the pump function
STOP_1 Stop the tempering function
STOP_4 Stop the pump function